Pace Mapping
If we pace from the exact spot where the tachycardia exits the myocardium, the QRS should match perfectly. The surface ECG becomes a homing beacon.
Every ventricular tachycardia has a fingerprint. The QRS morphology on the 12-lead ECG records the specific path the wavefront takes after it exits the reentrant circuit and spreads through the ventricular myocardium.
Pace mapping exploits a straightforward principle: if we place a pacing catheter at the exact exit site of that circuit and deliver a stimulus during sinus rhythm, we force the wavefront to spread through the ventricle along the same path. The result is a paced QRS that mirrors the clinical VT morphology.
A perfect 12-lead match tells us we are standing on the exit. A poor match tells us to move.
Why the Exit Determines the QRS
During VT, the electrical wavefront circulates through a protected channel within scar tissue. At the point where the wavefront exits this channel into healthy myocardium, it spreads outward freely. The direction and speed of that spread determine the QRS morphology we see on the surface ECG.
The critical insight: the QRS shape depends almost entirely on where the wavefront enters normal myocardium, and on the angle of that entry. Two circuits with different internal paths but the same exit site will produce nearly identical QRS complexes. The exit is the signature.
When we pace from that same exit site, we bypass the circuit entirely but recreate the identical activation sequence through the rest of the ventricle. If all 12 surface leads show a matching morphology, we have confirmed the location.
How to Score a Pace Map
We compare the paced QRS to the clinical VT QRS in all 12 leads, one by one. Each lead is scored as a match or a mismatch based on overall morphology, axis, and amplitude.
12/12 match: High confidence. The catheter is at the exit site.
10–11/12 match: Close. The catheter is within 1–2 cm of the exit. Minor differences in a single lead may reflect slight angulation or capture of adjacent tissue.
Below 10/12: The catheter is likely remote from the exit. Move and try again.
Scoring can be done visually at the bedside. Automated correlation algorithms such as PASO (Pace and Stop) or Barts correlation quantify the match mathematically, comparing the entire QRS waveform rather than relying on lead-by-lead visual assessment. These tools reduce subjective bias and improve reproducibility, particularly when differences are subtle.
Left: a 12/12 match confirms the catheter is at the VT exit site. Right: five mismatched leads indicate a remote pacing location.
The Stimulus-to-QRS Interval
When we pace from normal myocardium, the QRS begins almost immediately after the pacing stimulus. The stimulus-to-QRS interval (stim-to-QRS) is short, typically less than 40 ms.
When we pace from within scar tissue, the story changes. The electrical impulse must first travel through the slow-conducting channel inside the scar before it reaches healthy myocardium and generates a QRS on the surface. This produces a measurable delay between the stimulus artifact and the QRS onset.
A long stim-to-QRS interval tells us something important: the catheter is deep within the isthmus of the reentrant circuit, surrounded by scar. The wavefront has to traverse that protected corridor before it exits. We may have a perfect 12/12 QRS match, but the delay reveals that we are pacing from within the channel rather than from the exit itself.
This distinction matters for ablation planning. The exit site is one valid target. The protected isthmus is another. The stim-to-QRS interval helps us distinguish between them.
Pace Mapping vs Activation Mapping
Activation mapping requires the VT to be running. We move the catheter point by point, recording the local activation time at each location. The site with the earliest presystolic signal is closest to the origin. It is the gold standard for localizing the circuit, but it has a strict prerequisite: the VT must be sustained and hemodynamically tolerated long enough to complete the map.
Pace mapping works in sinus rhythm. We do not need the VT to be running at all. This makes it indispensable in two common scenarios: when VT is not inducible in the lab, and when VT causes hemodynamic collapse that forces us to terminate it immediately.
In practice, the two approaches complement each other. Activation mapping finds the earliest site during VT; pace mapping confirms the exit site from sinus rhythm. When both point to the same location, confidence is high. When we cannot perform activation mapping at all, pace mapping becomes the primary localization tool.
Limitations
Pace mapping has real constraints. The most common pitfall involves pacing output. Higher pacing output captures a larger volume of myocardium around the catheter tip, which can alter the QRS morphology slightly compared to the clinical VT. The general practice is to pace at the lowest output that achieves consistent capture, reducing the radius of captured tissue.
Epicardial VT exit sites pose a different problem. If the exit is on the outer surface of the heart, pacing from the endocardial surface may never replicate the morphology accurately. The wavefront would have to traverse the full wall thickness before exiting, producing a different activation pattern.
Patients with multiple VT morphologies require separate pace maps for each one. Each morphology represents a different exit site (and often a different circuit), so a single 12/12 match confirms only one of them.
Finally, pace mapping identifies the exit site, not the full isthmus. The protected channel through which the wavefront circulates during VT may extend several centimeters beyond the exit. A successful ablation often requires targeting the isthmus itself, which demands additional mapping strategies such as substrate mapping or entrainment.
Key Takeaways
- Principle: Pacing from the VT exit site reproduces the clinical QRS because the wavefront spreads through the ventricle along the same path.
- Scoring: A 12/12 lead match confirms the exit site. A 10–11/12 match places the catheter within 1–2 cm. Below 10/12 means the catheter is remote.
- Stim-to-QRS: A long stimulus-to-QRS interval indicates the catheter is deep within the scar channel, pacing through slow-conducting tissue before reaching the exit.
- Sinus rhythm advantage: Pace mapping works without sustained VT, making it essential when VT is non-inducible or hemodynamically unstable.
- Complementary tools: Activation mapping finds the earliest site during VT; pace mapping confirms the exit from sinus rhythm. Both pointing to the same location increases confidence.
- Constraints: Pacing output, epicardial exits, multiple VT morphologies, and the inability to map the full isthmus are the main limitations.